Explain the mechanism of each physical method of microbial control — flaming, incineration, pasteurisation, autoclaving, filtration, radiation, desiccation and cold — state the limit of each one, and justify why moist heat sterilises at 121 °C while dry heat needs 160–170 °C.
New to this course? How each section works
Learn gives you one short idea at a time. Each idea starts with a single sentence, then explains it, gives an example, and asks you a small question you can answer straight away.
Explore shows a diagram of the same ideas. Click the numbered parts to read what each one is.
Check asks a few questions. Answer them all correctly to finish the section and unlock the next one.
“Go deeper” is optional. It adds mechanism and extra detail for when you are ready — nothing you need for the checks is hidden there.
Why this matters
Almost every infection-prevention task you will perform rests on one of the methods in this section. Deciding whether an instrument goes into the autoclave or only needs a wipe, knowing that the pasteurised milk in the ward fridge is safe but not sterile, knowing that a filtered solution can still carry a virus or a toxin, and knowing that ultraviolet light cleans the air of one room but cannot reach inside a sealed package — these are clinical judgements, not laboratory trivia. A nurse who understands the mechanism can predict where a method fails before a patient is harmed by that failure; a nurse who has only memorised a list cannot.
What you need to know first
You need the vocabulary of the previous section — sterilisation, disinfection, asepsis — and nothing else. Five words are defined below: sterilisation, disinfection, bioburden, endospore and pathogen. The section is easier if you keep one question in mind for every method: what does it actually do to the cell, and what does it fail to do?
Words to know (5)
Sterilisation
A validated process that destroys or removes all forms of microbial life, including bacterial endospores, from an object or a fluid. Steam at 121 °C in an autoclave sterilises a surgical instrument pack.
Disinfection
A process that reduces the number of pathogenic microorganisms on a surface or an object, but which does not reliably destroy bacterial endospores. Boiling a thermometer disinfects it, but the instrument is not sterile afterwards.
Bioburden
The number and the resistance of the microorganisms present on an item or in a fluid before a control method is applied. A heavily soiled instrument needs a longer cycle than a clean one because its bioburden is higher.
Endospore
A dormant, highly resistant structure formed inside certain bacteria, able to survive heat, drying and radiation that would kill the growing cell. Clostridium and Bacillus species form endospores that survive boiling water.
Pathogen
A microorganism capable of causing disease in a susceptible host, which is the organism that pasteurisation and disinfection are chiefly aimed at. Salmonella in raw milk is a pathogen; the lactic acid bacteria of a yoghurt culture are not.
One idea at a timeIdea 1 / 7
In one sentence Physical control methods either destroy microorganisms outright or slow their multiplication until they can no longer spoil or infect.
Physical control removes or stops the growth of microbes
Microbial control can be achieved with chemicals, but this section deals with the physical agents: heat, radiation, filtration, drying and cold. Each agent is applied to a material — an instrument, a fluid, a food, a room — and each works because it attacks something the cell cannot do without. Heat denatures the cell’s proteins and destroys its membranes; radiation damages its genetic material; a filter physically withholds a particle larger than its pores; drying removes the water that the cell needs for metabolism; cold slows every enzyme reaction until growth almost stops. The choice of method is dictated by the object. Heat will ruin a plasma bag, radiation will damage many plastics, and a filter cannot be used on a solid.
For example
A metal wire loop is sterilised by flaming, whereas a plastic blood bag is sterilised during manufacture by gamma radiation and cannot be flamed.
Pause and check yourself
Why is the same method of control not used for every object that must be made safe?
Show the answer
Because each method attacks the cell in one particular way and each material tolerates only some of those attacks. Heat destroys proteins but also melts plastic; radiation damages DNA but also degrades some polymers; a filter withholds particles but cannot treat a solid. The method must suit both the organism and the object.
Read detailed explanations
How an autoclave is validated: temperature alone proves nothing
Setting the dial to 121 °C proves only that the chamber reached the set point somewhere; it does not prove that the steam penetrated the load, that the air was expelled, or that the organisms inside are dead. Validation therefore runs on three levels. Physical monitoring uses thermocouples threaded into the most difficult part of the pack, a pressure gauge and a chart or data logger that records the whole cycle, because the operator needs the temperature inside the load rather than in the jacket. Chemical indicators such as autoclave tape and internal strips change colour once the required temperature has been held, which confirms that the pack itself was exposed but says nothing about kill. Biological monitoring is the criterion: vials or strips carrying a defined population of Geobacillus stearothermophilus spores, an organism that is harmless to people and far more heat-resistant than Mycobacterium tuberculosis, are placed in the coldest, densest corner of the load and incubated afterwards. Failure of those spores to grow is the evidence of sterility; growth means the cycle failed even if every gauge read correctly. The reason a well-stocked laboratory does not rely on the gauge is the biphasic kill curve described by Gillespie and Ainsworth: a single exponential rate does not describe spore death, so the survival of a small resistant fraction cannot be excluded by temperature alone. Load configuration is the commonest cause of a false sense of security. A full chamber, an over-tight pack, a closed container, a load that is still wet at the end of the cycle or a residual air pocket all create a protected site where saturated steam never arrives and the exposure is shorter than intended. For that reason the routine checks are simple but not optional: the indicators are read, the printout is filed, the packs are stored dry and the shelf life is counted, and an unopened package may be judged sterile only while its wrapper is intact and dry. In hospital practice the nurse is the last quality-assurance step. Before an instrument is used on a patient, the pack should be inspected for a torn or damp wrapper, the chemical indicator should have changed as specified, and the batch record should be available; a load that failed biological monitoring is recalled and must never be released.
Why pasteurised milk is safe but not sterile
Pasteurisation is a public-health instrument aimed at named pathogens rather than at microbial life in general, and the temperatures were chosen to hit the most heat-resistant of those pathogens with a margin. The target that set the original holder method was Coxiella burnetii, the cause of Q fever, and the standard it produced is 63 °C for 30 minutes; the high-temperature short-time process that dairies actually use is 72 °C for 15 seconds, and ultra-high-temperature treatment, 135–150 °C for a few seconds, pushes the product into a different category altogether because it kills spores and allows months of unrefrigerated storage. The history matters because it explains the odd temperatures: Louis Pasteur and Claude Bernard showed in the 1860s that heating wine and beer to about 60 °C prevented souring without ruining the taste, and the process was named for him; the dairy industry later raised the temperature until the pathogen target was covered. What survives is what makes the milk non-sterile. Thermoduric bacteria such as enterococci, micrococci and some Bacillus species survive, and while they are not usually dangerous they multiply at refrigeration temperature and coagulate or sour the milk. Alkaline phosphatase, the enzyme assayed to check that pasteurisation was actually performed, is destroyed by the process and its absence is the test of adequacy, whereas heat-stable lipases and proteases from psychrotrophic bacteria are not destroyed and go on degrading the fat and the protein during cold storage, so a milk that was pasteurised correctly can still turn bitter. That is why the shelf life of pasteurised milk is measured in days under continuous refrigeration at about 4 °C and collapses rapidly above 8 °C, why a warm delivery van or a fridge door left ajar shortens it, and why the product must be dated and rotated. It is also why pasteurised milk is emphatically not a sterile product: it cannot be used for neonatal parenteral nutrition, it is not interchangeable with UHT milk, and it cannot be left at room temperature as though it were canned.
Dry heat, moist heat and the arithmetic of killing
The gap between 121 °C and 170 °C is not an accident of equipment; it follows from how heat kills and from what each form of heat does to the cell. Moist heat kills chiefly by coagulating proteins, and water is an excellent conductor, so saturated steam condenses on the cooler load, releases latent heat and drives the surface to the chamber temperature almost immediately. Dry heat acts much more slowly: it kills partly by oxidising cell constituents rather than by coagulation, and air conducts poorly, so the load must be held at a higher temperature for a long time to accumulate the same damage. This is why the standard moist cycle is 121 °C for 15 minutes at about 15 psi, or 134 °C for 3 minutes at about 30 psi, while the standard dry-heat reference is 160 °C for 2 hours or 170 °C for 1 hour. The supplied notes give dry heat as 170 °C for 120 minutes on lines 1736–1737; the correct standard pair is 160 °C for two hours or 170 °C for one hour, and the notes’ combination overstates the time needed at 170 °C. Moist heat also penetrates more predictably, which is why it is preferred for fabrics, liquids, media and instruments, while dry heat is reserved for glassware, metal, oils, powders and sharp items that steam cannot reach or that corrosion or moisture would damage. Behind both methods sits one quantitative rule: microbial killing by heat is approximately exponential, so the population falls by a constant proportion per unit time rather than by a constant number. The D-value, the decimal reduction time, is the time at a stated temperature needed to reduce the population by 90 %, or one log; a sterilisation cycle is then designed to deliver enough log reductions to bring an initial bioburden down to a defined probability of a surviving organism, conventionally one in a million for a sterilised item. Two consequences follow. First, the cycle time depends on the bioburden and on how resistant the contaminating spores are, so heavily soiled instruments require longer, and organic matter such as blood or pus must be removed before sterilisation because it shields organisms and can even protect them chemically. Second, temperature and time are interchangeable to a degree: a higher temperature with a shorter holding time can achieve the same log reduction, which is the whole logic of flash cycles. For the nurse this arithmetic explains two habits that otherwise look like fussiness: never assume that a short cycle at the right temperature is automatically enough, and never place a cold or wet instrument into the chamber, because the moisture left on the surface will delay the rise of the whole load to the required temperature.
Explore the diagram
Look at the whole drawing first. Then select one numbered part and connect it to the idea you just read.
The physical methods of microbial control: what each one does, and where each one stops
Every method attacks one thing the cell cannot do without, and every method has a blind spot. Follow the bands from top to bottom: the flame and the incinerator burn; pasteurisation and the autoclave use heat, one below boiling and one as saturated steam; filtration traps; radiation either penetrates or merely touches the surface; drying and cold only slow. The third line of each band names the limit — the fact that decides whether the object in your hand is sterile, disinfected or merely safe.
New to diagrams? Read this first
Look at the whole drawing once without clicking anything. Ask yourself only: what kind of picture is this (a comparison, a route, a sequence, a table)?
Press Play, or use the arrows, to move through the numbered parts in order. The drawing stays still; only the highlight moves.
Click a number, or its name in the list below, to read what that part is and why it matters. Use “Hide labels” when you want to test your memory.
More controls
01 / 6Heat that burns: flaming and incineration
If the drawing is wider than this panel, scroll it sideways. Labels stay readable; the numbered explanations below show the same ideas.
Heat that burns: flaming and incineration
Flaming holds a metal object in the flame for a few minutes, and the flame burns out the microbes directly; it is how loops, needles and tube mouths are sterilised between transfers. Incineration is not an instrument method at all but a waste-treatment process: it combusts the organic fraction of waste and converts it into ash, flue gas and heat, so nothing survives to be stored or transported.
Comparative map, not to scale: the temperatures and times are the standard reference values for each method and vary with the load, the organism and the equipment. Validated sterilising methods include saturated steam, dry heat at 160 °C for 2 hours or 170 °C for 1 hour, flaming of metal items, ionising radiation and low-temperature chemical sterilants such as ethylene oxide; among the methods shown, only saturated steam and ionising radiation sterilise a packaged load. Ultraviolet light and membrane filtration do not sterilise.
Key points to remember
✓Physical control kills or stops microbes, and the method chosen must suit both the organism and the material.
✓Flaming burns microbes directly on metal; incineration combusts waste into ash, flue gas and heat and destroys it entirely.
✓Pasteurisation holds 63 °C for 30 minutes or 72 °C for 15 seconds, kills pathogens and inactivates spoilage enzymes, but leaves spores alive.
✓Autoclaving sterilises with saturated steam at 121 °C because steam carries latent heat; dry heat needs 160 °C for two hours or 170 °C for one hour.
✓Filtration traps bacteria and larger particles only, so viruses, mycoplasmas and toxins pass through and the filtrate is not truly sterile.
✓Ionising radiation penetrates and damages the cell; non-ionising ultraviolet causes thymine dimers but cannot pass through packaging or liquids.
✓Desiccation and cold are reversible: both control growth, neither sterilises, and psychrophiles still grow in the cold.
Go deeper
Worked example
Choosing a control method in the clinic. You are helping to prepare a minor-surgery room. Three items arrive on the bench at the same time: a stainless-steel needle-holder used to suture a laceration, a glass bottle of a heat-sensitive drug solution that must be given intravenously, and a box of sealed single-use syringes. For each item you must decide which control method is appropriate, what it achieves, and what it does not achieve.
Classify each item by contact with the patient. The needle-holder will pass through skin and tissue, so it is a critical item and must be sterile. The drug solution will enter the bloodstream directly, so it must be free of living organisms. The sealed syringes never touch the patient before use, so they only need to reach the room intact and sterile as supplied.
Send the metal instrument to the autoclave with its cycle in mind. Steam at 121 °C for 15 minutes at about 15 psi, or 134 °C for 3 minutes, will destroy endospores provided that the air has been removed and the steam has reached the whole load. Do not boil it: boiling at 100 °C leaves endospores alive.
Filter the heat-sensitive solution instead of heating it. A 0.22 µm membrane filter withholds bacteria and larger particles, so the drug keeps its activity and the bacteria are removed. Remember the limit: viruses, mycoplasmas and any toxins present pass through, so the filtrate is not sterile in the full sense and must still be handled as a clean pharmaceutical product with its own expiry and storage rules.
Leave the sealed syringes sealed, and check the other two items afterwards. Single-use items are usually sterilised during manufacture by ionising radiation or by ethylene oxide, and sterility holds only while the packaging is intact; ultraviolet light would not reach inside the pack. Once the autoclave cycle is finished, read the indicator and the cycle record before you use the instrument.
One room, three objects, three different methods — and each choice follows from the same two questions: what does the object have to be, and what does the method do to the cell? The instrument had to be sterile, so it needed steam under pressure. The solution had to keep its activity, so it needed filtration and accepted the loss of sterility. The syringes were already sterile, so the only task left was to keep the packaging intact.
Common misconception
The commonest beginner error is to treat every method in this section as though it sterilised, and therefore to use “sterile”, “disinfected” and “clean” as if they were interchangeable. Boiling is the trap: the water is at 100 °C and visibly furious, yet bacterial endospores survive it, so a boiled instrument is disinfected and not sterile, and an instrument that will enter tissue must go to the autoclave. A second error mirrors the first: assuming that because a method is powerful it is also complete. Gamma radiation sterilises a sealed pack, but ultraviolet light of the same family cannot reach through the pack at all; a membrane filter removes bacteria but lets viruses, mycoplasmas and toxins through. The reliable habit is to name the mechanism and then name its blind spot.
Check your reasoning
Choose one object you handle in a clinical setting — a dressing trolley, a stethoscope, a laryngoscope blade, a milk feed, a specimen container. Which method of control is actually applied to it, what does that method achieve, and what does it leave untouched?
Reveal an answer
Take a stethoscope used between patients. It is cleaned with detergent and an alcohol wipe, which is degerming and disinfection: it removes most of the transient organisms from the surface of the diaphragm and the tubing, and it is entirely appropriate because the stethoscope does not breach skin. What it leaves untouched is the deep crevice in the earpieces and, more importantly, the resistant spores and any virus in a protected crack, and the method is only as good as the friction and the contact time. Nothing about the stethoscope is sterilised, and nothing needs to be.
Where this comes from
BMS 114 General Microbiology I lecture notes (supplied) · “Flaming … Refrigeration and Freezing”, lines 1764–1802: flaming, incineration, pasteurisation, autoclaving, filtration, radiation, desiccation, refrigeration and freezing.
Correction to the supplied notes — pasteurisation temperatures · Correction to the supplied notes: lines 385–386 give Pasteur’s original wine treatment as 50–60 °C for a few minutes. That is a historical observation, not today’s dairy standard, which is 63 °C for 30 minutes (holder) or 72 °C for 15 seconds (high-temperature short-time); both are below boiling and neither sterilises.
Correction to the supplied notes — filtration is not sterilisation · Correction to the supplied notes: lines 1783–1785 state that the fluid received from a filter “has no bacteria as it is sterilized”. A filter withholds bacteria and larger particles but not viruses, mycoplasmas or toxins, so the filtrate is not sterile in the full sense and the notes’ claim is too strong.
Correction to the supplied notes — dry-heat values · Correction to the supplied notes: lines 1736–1737 give dry heat as 170 °C for 120 minutes. The standard reference pair is 160 °C for 2 hours or 170 °C for 1 hour; the notes’ combination overstates the time needed at 170 °C.
Beyond the supplied notes — autoclave cycle, validation and the log-kill rule · the notes at lines 1781–1782 name autoclaving and its effect on spores without giving the cycle, the three levels of validation (physical, chemical, biological with Geobacillus stearothermophilus), load configuration, the D-value and the biphasic spore-kill curve of Gillespie and Ainsworth follow standard sterilisation practice and microbiology; the notes’ pasteurisation paragraph (lines 1776–1780) is silent on which pathogens set the temperature.
Beyond the supplied notes — ionising versus non-ionising radiation and the limits of drying and cold · the notes at lines 1786–1791 distinguish ionising from non-ionising radiation but do not state what the penetration difference means in practice; the consequences for air and surface decontamination, for packaged items, for the reversibility of desiccation and for the growth of psychrophiles in refrigerated food are standard microbiology.
Lesson progress is saved on this device for the current learner. Practice answers stay separate and are not stored.
About this course and its sources
Bilingual lessons organised from the supplied BMS 114 General Microbiology I lecture notes (Mr Ngotelo F. Ambene, Mr Lyonga B. Lyonga & Mr Fumbui Louis Ngong). Diagrams are original educational schematics; sections that correct the notes or go beyond them say so in their sources. Nursing and French-language review are pending.